Temperature

Setting temperature range-10 to +60 °C
Working temperature rangefrom -10°C up to +60°C without humidity (not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over)
Working temperature rangefrom +10°C up to +60°C with humidity and/or light
Setting accuracy temperature0.1 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity

Humidityhumidity supply with distilled water from external tank by self-priming pump
Humidificationhumidification by hot steam generator
Humidityactive humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity – resolution of display 0.1 %, setting accuracy 0.5 %

Control technology

ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
adjustable parameterstemperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable values each, temperature/humidity, for two freely selectable values CO2

Ventilation

Fanforced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps

Communication

Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature controlmechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature controlover- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETYadditionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic systemintegral fault diagnostics for temperature and humidity control
Alarmvisual and acoustic

Heating concept

Air jacketair jacket heating system for gentle all-around heating
Heating conceptno drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Coolingenergy-saving use of CFC-free cooling/heating system (refrigerant R290 Propan)
Defrostinghighly efficient automatic defrosting system

Standard equipment

Works calibration certificatestandard values +10 °C and +37 °C, 60 % rh at +30 °C
Doorinner glass doors
Doorfully insulated stainless steel doors with 2-point locking (compression door lock)
Entry portEntry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back
Internals2 stainless steel grid(s), electropolished
Standard accessoriesWater tank including connection hose

Stainless steel interior

Dimensionsw(A) x h(B) x d(C): 1040 x 1200 x 600 mm (d less 33 mm for fan)
Volume749 l
Max. number of internals14
Max. loading of chamber200 kg
Max. loading per internal30 kg

Textured stainless steel casing

Dimensionsw(D) x h(E) x d(F): 1224 x 1950 x 785 mm (d +56mm door handle)
Installationon lockable castors
Housingrear zinc-plated steel

Electrical data

Voltage Electrical load230 V, 50/60 Hz
approx. 1350 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1350 W

Ambient conditions

Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature15 °C to 28 °C (up to 34 °C with limited temperature range)
Humidity rhmax. 70 %, non-condensing
Altitude of installationmax. 2,000 m above sea level
Overvoltage categoryII
Pollution degree2

Packing/shipping data

Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 1330 x 2150 x 1050 mm
Net weightapprox. 249 kg
Gross weight cartonapprox. 319 kg

ICH750

Standard delivery Climate chamber

On this page, you can find all the essential technical data on the Memmert climate chamber ICH. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists

Ulasan

Belum ada ulasan.

Jadilah yang pertama memberikan ulasan “Climate Chamber ICH750 Memmert”

Alamat email Anda tidak akan dipublikasikan. Ruas yang wajib ditandai *

2
Rp57.000.000

Cart